Throughput
Flexibility
Reagents & Samples
Small Sample Ready
Photometry Range
As part of Mindray Animal Medical's Life Science portfolio, the respons® 910VET supports clinical chemistry testing in academic institutions, research and reference laboratories, and preclinical and clinical research settings. The system performs routine, specialty, and calculated chemistry testing on a single platform, helping laboratories generate reproducible results and high-quality research data.
| Parameter | Target | Mean | Recovery % | CV % Intra | CV % Inter | |
|
AMY (U/L) |
N | 74.7 | 75.8 | 102 | 1.64 | 2.63 |
| P | 250 | 258 | 103 | 1.8 | 2.12 | |
|
CHOL (mg/dL) |
N | 135 | 140 | 104 | 2.13 | 2.29 |
| P | 204 | 210 | 103 | 1.66 | 2.86 | |
|
GGT (U/L) |
N | 28.3 | 29.2 | 103 | 1.77 | 1.70 |
| P | 89.5 | 89.4 | 99.9 | 1.92 | 1.48 | |
|
TP (g/dL) |
N | 5.28 | 5.18 | 98 | 1.22 | 1.39 |
| P | 6.45 | 6.5 | 101 | 0.94 | 1.13 | |
|
UREA (mg/dL) |
N | 43.9 | 44.9 | 102 | 2.48 | 3.58 |
| P | 156 | 170 | 109 | 2.11 | 2.28 |
N-TruLab N = Normal control; P-TruLab P = Pathological control n=20
Innovation is a core principle upon which we manage our Life Science business. Our instrument line continues to expand with higher-throughput systems designed to deliver comprehensive menu options in higher-volume settings. Our offerings are growing across clinical chemistry, hematology, electrolyte testing, cell morphology, imaging, patient monitoring, life support, and integrated research laboratory solutions. This commitment to innovation keeps our customers at the forefront of the latest technological advances in research laboratory diagnostics. All systems are designed to provide flexibility, allowing laboratories to configure testing workflows according to their clinical, operational, and budgetary requirements.
Reagent applications for the respons910®VET system are specifically configured to minimize the impact of interfering substances such as lipids, bilirubin and hemoglobin in samples. Extended linearity of our assays nearly eliminates repeat testing due to sample results being out of range. Our fully integrated system—made up of easy-to-use analyzer, proprietary reagents and optimized applications—assures the most cost-effective and highest-quality results, time after time. Quality you can count on. To us, the proof is in the results!
| Technical Information | |
| System type | Bench top clinical chemistry analyzer |
| Throughput | 150 tests/hour |
| Reagent/sample tray | 30 reagent positions plus 30 sample positions; combined in single tray |
| Dead volume | As little as 5 µL |
| Sample types | Serum, plasma, whole blood, urine, and other fluids |
| Sample volume | 2 – 30 μL |
| Reagent pipetting volume | Reagent 1: 120 – 250 μL Reagent 2: 10 – 130 μL |
| Sensors | Liquid-level sensor, clot sensor and crash sensor |
| STAT-analytics | 2 sample positions for loading of emergency samples at any time |
| Bar code identification | Automatic bar code scan for reagents and samples |
| Measuring principle | Colorimetry (rate; end point) |
| Sample tubes/cups | Primary tubes of 3, 5, 7 and 10 mL and sample cups and microcentrifuge tubes (0.5, 1.5 and 2.5 mL) |
| Reaction temperature | 37 ± 0.2 °C |
| Photometry | 12 wavelengths: 340, 380, 405, 450, 480, 508, 546, 570, 600, 660, 700 and 800 nm (mono- and bichromatic) |
| PMS/LIS connectivity | Yes |
| Remote diagnostics | Yes |
| System interface | Analyzer - PC: USB connectivity bi-directional; CPU: Pentium IV or higher |
| Power source | AC 110/220 V, 60/50 Hz; 300 VA excluding PC/printer/monitor |
| Dimensions | 24 in (W) x 26 in (D) x 24 in (H) / 60 cm (W) x 67 cm (D) x 60 cm (H) |
| Weight | 132 lbs / 60 kg |